Nuclear Demagnetization Systems are used cool samples to ultra-low temperatures (below 4mK).
Cryomagnetics' Nuclear Demagnetization Magnets incorporate a main high field coil with series wound cancellation coils. The location and strength of the cancellation coils are specified by the customer.
These magnets are also commonly used with dilution
refrigerators (< 10mK).
Cryomagnetics has manufactured NbTi Nuclear Demagnetization Magnets up to 9T (4.2K), 11T (2.2K). Nb3Sn based versions delivered include up to 17T (4.2K), 19T (2.2K).
Pictured at left is a 14T/15T
(4.2K/2.2K) high field magnet with compensation coils.
Click on the "Specifications" tab above for details on
this superconducting magnet.
At right is a 13T/14T (4.2K/2.2K)
high field magnet with compensation coils.
Contact us with your specifications.

Example High Field
Superconducting Magnet with Compensation Coils
Use this example to help define your specific
needs...
Superconducting Magnet, Model
140/150-200-010D
Solenoid configuration.
14 Tesla central field (4.2K).
15 Tesla central field (2.2K).
+ 0.1% central field homogeneity over a 10 mm DSV.
Compensation region of + 40 gauss from 26 cm to 36
cm on-axis measured from high field magnetic center.
2.00 inches clear bore.
Persistent mode switch installed.
Fully protected against damage due to quench.
Please click on the following
links for more information:
Please Contact Cryomagnetics at sales@cryomagnetics.com or your local Sales Representative for a price quotation.